Ventec Life Systems recalls VOCSN+Pro and V+Pro ventilators with incorrect parts that could cause inaccurate tidal volume monitoring and unexpected shutdowns. Affected units have specific serial numbers and model references.
Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.
Ventec Life Systems is recalling VOCSN+Pro and V+Pro multi-function ventilators because they were serviced with incorrect parts. This could lead to inaccurate breathing monitoring, potentially causing the ventilator to shut down unexpectedly during use.
Incorrect parts used during servicing can cause the ventilator to shut down unexpectedly or provide inaccurate breathing measurements. This may result in less inspiratory volume, pressure, or PEEP delivery, which could compromise patient safety during critical medical procedures.
Healthcare professionals using Ventec Life Systems' VOCSN+Pro or V+Pro ventilators are most at risk. Patients relying on these ventilators for breathing support could be affected if the device malfunctions.
Check for the model names VOCSN+Pro (V+O+C+S+N+Pro) or V+Pro (V+Pro) with specific serial numbers listed in the recall notice. The reference numbers include PRT-00490-001 and PRT-01185-000/002.
Identify if your ventilator is a VOCSN+Pro or V+Pro model with the specific reference numbers listed in the recall.
Reach out to Ventec Life Systems directly for guidance on whether your unit is affected and next steps.
The ventilators were serviced using incorrect parts, which can cause unexpected shutdowns or inaccurate breathing monitoring.
The ventilator might shut down unexpectedly or deliver incorrect breathing support, potentially leading to less inspiratory volume, pressure, or PEEP.
Check the serial numbers against the list provided in the recall notice, which includes specific reference numbers like PRT-00490-001 and PRT-01185-000/002.
We’ve drafted a message you can send Ventec Life Systems to request your refund or repair — edit it as you like.
Subject: Recall Z-1196-2025 — Ventec Life Systems Ventec Life Systems Hello, I own a Ventec Life Systems that is covered by recall Z-1196-2025 from Ventec Life Systems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.