Baxter Healthcare recalls TruSystem 7500 U mobile column with software issue that makes upper back section unadjustable when emergency mode is enabled. Affected units have specific serial numbers and were sold until November 2024.
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Baxter Healthcare is recalling its TruSystem 7500 U mobile column because a software problem causes the upper back section to become unadjustable when emergency mode is enabled. This could prevent proper adjustment during critical medical situations, potentially affecting patient safety.
The software issue causes the upper back section to become unadjustable when emergency mode is enabled. This could prevent proper adjustment of the column during critical medical situations, potentially affecting patient safety. The hazard occurs only when emergency mode is activated, which may happen during specific medical procedures.
Healthcare facilities using the TruSystem 7500 U mobile column are most at risk. Patients and staff could be affected if the column becomes unadjustable during emergency mode activation.
Check for the product code 1730720 and serial numbers manufactured until November 14, 2024. The affected units have specific UID/DI numbers starting with 00887761974388.
Verify if the upper back section remains adjustable when emergency mode is enabled.
The software issue causes the upper back section to become unadjustable when emergency mode is enabled, which could affect patient safety during critical medical situations.
Check for product code 1730720 and serial numbers manufactured until November 14, 2024. Affected units have specific UID/DI numbers starting with 00887761974388.
The recall does not specify a remedy, so check if the upper back section remains adjustable when emergency mode is enabled.
We’ve drafted a message you can send Baxter Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-1413-2025 — Baxter Healthcare Baxter Healthcare Hello, I own a Baxter Healthcare that is covered by recall Z-1413-2025 from Baxter Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.