GE Healthcare recalls specific CARESTATION 750c A1 anesthesia systems (model 1012-9755-000) due to potential ventilation failure in Volume Control Ventilation mode. Affected units may require repair or replacement.
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
GE Healthcare is recalling certain CARESTATION 750c A1 anesthesia systems that may not provide effective ventilation in Volume Control Ventilation (VCV) mode. This could lead to inadequate breathing support during surgery if the system fails to function properly.
The anesthesia delivery system may fail to provide adequate ventilation when set to Volume Control Ventilation mode. This could result in insufficient oxygen delivery to patients during surgery. The system can still function safely in Pressure Control Ventilation or Manual ventilation modes.
Healthcare providers using the specific model and serial numbers listed in the recall notice are most at risk. Patients undergoing anesthesia with these systems could face breathing complications if the device malfunctions.
Check for the model number 1012-9755-000 on the device or its packaging. Affected units have specific serial numbers listed in the recall notice, including SQT23330001WA through SQT24440014WA.
Verify the model number on the device or packaging to confirm if it matches 1012-9755-000.
Reach out to GE Healthcare for guidance on whether your unit requires repair or replacement.
The anesthesia system may not provide effective ventilation in Volume Control Ventilation mode, which could lead to inadequate breathing support during surgery.
Check the model number and serial numbers listed in the recall notice. Contact GE Healthcare for repair or replacement instructions.
The system can still function safely in Pressure Control Ventilation or Manual ventilation modes, so risks are manageable with proper use.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-1637-2025 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-1637-2025 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.