Radiometer ABL90 FLEX analyzers with Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze during sample measurements, causing test delays. Affected units include specific model numbers and serials listed in the recall.
ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.
Radiometer ABL90 FLEX and ABL90 FLEX PLUS analyzers with Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may freeze during sample measurements. Even after restarting, samples being processed at the time will be lost, potentially delaying test results. This recall affects specific model numbers and serials.
The analyzer software includes a watchdog feature that detects freezes and automatically restarts the software. However, any samples being processed during a freeze are lost, which can delay test results. This does not cause injury but may lead to delays in critical medical testing.
Healthcare professionals using these analyzers in labs, near patients, or point-of-care settings are most at risk. The recall affects specific model numbers and serials listed in the notice.
Check for model numbers 393-090 or 393-092 on the device. Affected units have specific serial numbers listed in the recall notice, including those starting with 090R or 092R.
Look for model numbers 393-090 or 393-092 on the analyzer.
Confirm the serial number is listed in the provided recall list (e.g., 090R0643N002, 092R0150N001).
No, but you should check if your device is affected by this recall. The recall does not cause injury but may delay test results if the analyzer freezes.
The analyzer automatically restarts, but any samples being processed at that time will be lost. This is a known issue with specific software versions.
Check the model number (393-090 or 393-092) and serial number against the list provided in the recall notice.
We’ve drafted a message you can send Radiometer to request your refund or repair — edit it as you like.
Subject: Recall Z-1659-2025 — Radiometer Radiometer Hello, I own a Radiometer that is covered by recall Z-1659-2025 from Radiometer. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.