Abbott Diagnostics recalls Architect Total T36X4ML calibrators due to transit delays from thunderstorms causing improper storage. This may lead to incorrect or delayed test results. Affected units have specific catalog and GTIN numbers.
transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an incorrect test result or a delayed test result.
Abbott Diagnostics is recalling its Architect Total T3:6X4ML calibrators because severe thunderstorms caused some units to be stored outside recommended conditions. This could lead to incorrect or delayed test results if used.
Severe thunderstorms caused transit delays for some calibrators, leading to improper storage conditions. This could result in incorrect test results or delayed testing if the device is used without proper calibration.
Healthcare facilities using the Abbott Architect Total T36X4ML calibrator are likely affected. Those handling diagnostic tests with this device are at risk of inaccurate results.
Check for the catalog number 07K6402 or GTIN 380740144845 on the calibrator. The product was sold with order number 46061983.
Ensure the calibrator is stored within the recommended temperature and humidity conditions to prevent incorrect test results.
Severe thunderstorms caused transit delays for some units, leading to improper storage conditions that could increase the risk of incorrect or delayed test results.
Check the catalog number 07K6402 or GTIN 380740144845 to confirm if your unit is affected. No specific action is required beyond ensuring proper storage conditions.
The recall does not provide a fix; users must ensure proper storage conditions to avoid incorrect test results.
We’ve drafted a message you can send Abbott Diagnostics to request your refund or repair — edit it as you like.
Subject: Recall Z-1849-2025 — Abbott Diagnostics Abbott Diagnostics Hello, I own a Abbott Diagnostics that is covered by recall Z-1849-2025 from Abbott Diagnostics. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.