Recalled SAFE-T-FILL micro capillary blood collection tubes (July 31, 2023 - Feb 28, 2025) cause false positive lead results when used with Magellan Diagnostics LeadCare Testing Systems, leading to unnecessary testing.
All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Anatomy Supply Partners is recalling certain SAFE-T-FILL micro capillary blood collection tubes that can cause false positive lead results when used with Magellan Diagnostics LeadCare Testing Systems. This may lead to unnecessary testing and delayed results for patients.
These blood collection tubes can produce false positive lead results when used with Magellan Diagnostics LeadCare Testing Systems. False positives may cause unnecessary follow-up testing and delays in getting accurate results.
People who use these blood collection tubes with Magellan Diagnostics LeadCare Testing Systems. Those most at risk are patients who have had blood tests using these specific tubes.
Check for tubes labeled 'SAFE-T-FILL Micro Capillary Blood Collection' with model numbers 07 6011, 07 6013, 07 7051, 07 7052, 07 7053, 07 7056, or 07 7058. These were manufactured between July 31, 2023 and February 28, 2025.
Look for model numbers 07 6011, 07 6013, 07 7051, 07 7052, 07 7053, 07 7056, or 07 7058 on the tubes.
These tubes can cause false positive lead results when used with Magellan Diagnostics LeadCare Testing Systems, which may lead to unnecessary testing and delayed results.
Check if your tubes have the model numbers listed above. If they do, stop using them with Magellan Diagnostics LeadCare Testing Systems and contact the manufacturer for guidance.
The recall is for false positive results, not physical safety risks. There is no risk of injury from the tubes themselves.
We’ve drafted a message you can send Anatomy Supply Partners to request your refund or repair — edit it as you like.
Subject: Recall Z-0372-2026 — Anatomy Supply Partners Anatomy Supply Partners Hello, I own a Anatomy Supply Partners that is covered by recall Z-0372-2026 from Anatomy Supply Partners. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.