Bridge to Life recalls EasiSlush (0.9% Sodium Chloride Solution) due to uncompleted quarterly dose audits in early 2024. This affects specific lots for sterile slush preparation in medical settings.
The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024. Routine lot by lot testing was completed properly however the quarterly dose audit (sampling/testing to monitor the gamma sterilization process and confirm the validated parameters remain effective to assure sterility and bioburden reduction per requirements) was not completed on time.
Bridge to Life is recalling specific lots of EasiSlush (0.9% Sodium Chloride Solution) because the company discovered three lots failed to complete required quarterly dose audits in early 2024. This could lead to improper sterilization if the solution is used in medical procedures without proper validation.
The quarterly dose audit checks the gamma sterilization process to ensure the solution remains sterile and effective. Failure to complete these audits means the solution might not meet sterility standards, potentially causing infections during medical procedures.
Healthcare facilities using the recalled lots of EasiSlush for sterile slush preparation in medical procedures. Patients and staff at these facilities are at risk if the solution is used without proper validation of the sterilization process.
Check the lot number on the product label: 240372708, 240402836, or 240402837. The product is a sterile sodium chloride solution used for medical slush preparation.
Look for lot numbers 240372708, 240402836, or 240402837 on the product label.
Bridge to Life discovered three lots of EasiSlush failed to complete required quarterly dose audits in early 2024, which could affect sterility validation.
The risk is that the solution might not meet sterility standards if the gamma sterilization process is not properly validated, potentially leading to infections during medical procedures.
Check the lot number on the product label: 240372708, 240402836, or 240402837.
We’ve drafted a message you can send Bridge to Life to request your refund or repair — edit it as you like.
Subject: Recall Z-2157-2025 — Bridge to Life Bridge to Life Hello, I own a Bridge to Life that is covered by recall Z-2157-2025 from Bridge to Life. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.