Draeger recalls SafeStar 55 Plus filters due to potential CO2 curve increases during gas sampling. Affected units include all lots of this bidirectional breathing system filter.
Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters.
Draeger is recalling all SafeStar 55 Plus filters because they may cause a slow increase in CO2 levels when gas is sampled from the filter's port. This could lead to inaccurate readings in breathing systems.
The filter may cause a slow increase in CO2 levels when gas is sampled from its port. This could lead to inaccurate readings in the breathing system, potentially affecting medical monitoring.
Users of the SafeStar 55 Plus filter for bidirectional breathing systems. Those with medical devices requiring precise gas monitoring are most at risk.
Check for the SafeStar 55 Plus filter with model number MP05790. All lots are affected, including those sold with UDI 04048675558736.
Identify the filter using the model number MP05790 or UDI 04048675558736.
The filter may cause a slow increase in CO2 levels during gas sampling, which could lead to inaccurate readings in breathing systems.
Check for the SafeStar 55 Plus filter with model number MP05790 or UDI 04048675558736.
The recall does not specify a remedy, so contact Draeger for guidance.
We’ve drafted a message you can send Draeger to request your refund or repair — edit it as you like.
Subject: Recall Z-2159-2025 — Draeger Draeger Hello, I own a Draeger that is covered by recall Z-2159-2025 from Draeger. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.