GE Medical Systems recalls SIGNA Artist Evo MRI systems where caster locks weren't properly checked during maintenance since March 2023, potentially causing patient table instability.
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
GE is recalling SIGNA Artist Evo MRI systems that may not have had caster locks checked during maintenance since March 2023. This could lead to patient table instability during use, which might cause injury.
The caster locks on the MRI patient table may not have been properly checked during planned maintenance. This could cause the table to become unstable during use, potentially leading to patient injury.
Hospitals and imaging centers that own SIGNA Artist Evo MRI systems. Patients using these systems are at risk if the table becomes unstable during scans.
Check for system IDs that GE HealthCare performed maintenance on since March 1, 2023. Look for GTIN numbers: 00195278481382, 00840682118118, 00840682123129, 00840682147095.
Review system maintenance logs to confirm if GE HealthCare performed planned maintenance since March 1, 2023.
GE HealthCare has found that for certain MRI systems, the planned maintenance steps to check the caster locks on the patient table may not have been performed.
Check your system maintenance records to see if GE HealthCare performed planned maintenance since March 1, 2023. If not, contact GE Medical Systems for guidance.
Yes, if the patient table becomes unstable during use, it could lead to patient injury.
We’ve drafted a message you can send GE Medical Systems to request your refund or repair — edit it as you like.
Subject: Recall Z-2310-2025 — GE Medical Systems GE Medical Systems Hello, I own a GE Medical Systems that is covered by recall Z-2310-2025 from GE Medical Systems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.