Beckman Coulter recalls UniCel DxI 800 Access Immunoassay Analyzers with linear slide PSL software version 1.0.0.1 or earlier. These units may have incorrect motor settings that cause test delays. Affected units have serial numbers starting with 610... or 630...
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
Beckman Coulter recalls UniCel DxI 800 Access Immunoassay Analyzers with linear slide PSL software version 1.0.0.1 or earlier. These units may have incorrect motor settings that cause test delays. Affected units have serial numbers starting with 610... or 630...
Incorrect motor settings may cause test delays
Users of UniCel DxI 800 Access Immunoassay Analyzers with linear slide PSL software version 1.0.0.1 or earlier
Serial numbers starting with 610... or 630...
Affected units have serial numbers starting with 610... or 630...
Software version 1.0.0.1 or earlier
Look for serial numbers starting with 610... or 630...
We’ve drafted a message you can send Beckman Coulter to request your refund or repair — edit it as you like.
Subject: Recall Z-2266-2025 — Beckman Coulter Beckman Coulter Hello, I own a Beckman Coulter that is covered by recall Z-2266-2025 from Beckman Coulter. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.