Philips recalls Achieva 3.0T MRI systems with specific model numbers due to potential fire risk from gradient coil failures. Affected units may produce smoke or fire if components fail.
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
Philips is recalling certain Achieva 3.0T MRI systems because a component in the gradient coil could fail and cause smoke or fire. This is serious enough that users should stop using the affected models immediately.
The gradient coil in the affected MRI systems can fail and act as a heat source, potentially producing smoke or fire. This hazard occurs during normal operation and could lead to serious injury or property damage.
Medical facilities that own or use the affected Achieva 3.0T MRI systems are most at risk. These systems are used in hospitals and clinics for medical imaging.
Check the model numbers listed: 781177, 781277, 781278, 781344, or 781345. Also verify the serial numbers provided in the recall notice.
Immediately stop using the affected Achieva 3.0T MRI systems to prevent potential fire or smoke production.
Model numbers 781177, 781277, 781278, 781344, and 781345 are affected.
Check the model numbers listed in the recall: 781177, 781277, 781278, 781344, or 781345. Also verify the serial numbers provided in the recall notice.
The recall does not specify a fix; the remedy is null, meaning users must stop using the affected systems immediately.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-2200-2025 — Philips Philips Hello, I own a Philips that is covered by recall Z-2200-2025 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.