MAQUET recalls Corin Operating Tables with specific model numbers and software versions due to potential interference with high-frequency surgical equipment, causing unresponsiveness.
Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.
MAQUET is recalling Corin Operating Tables with model numbers 770001A0 through 770001F2. The tables may become unresponsive when used with high-frequency surgical equipment, which could disrupt medical procedures.
The operating table can become unresponsive to commands from both the universal remote and override panel when used simultaneously with high-frequency surgical equipment. This interference may prevent the table from functioning properly during critical medical procedures.
Healthcare professionals using these operating tables with high-frequency surgical equipment are most at risk. Patients and staff in medical facilities that use the recalled models may experience disruptions during procedures.
Check for model numbers 770001A0, 770001A2, 770001B0, 770001B2, 770001F0, or 770001F2. These tables were sold with all software versions and specific serial numbers listed in the recall notice.
Identify your operating table's model number to see if it matches the recalled models (770001A0 through 770001F2).
Ensure your table's software version is included in the recall, as all versions are affected.
The operating table may become unresponsive when used with high-frequency surgical equipment, which could disrupt medical procedures.
Check your model number against the list of recalled models (770001A0 through 770001F2) and software versions. No specific remedy is provided in this recall notice.
The recall notice does not specify a fix or remedy for the issue.
We’ve drafted a message you can send MAQUET to request your refund or repair — edit it as you like.
Subject: Recall Z-2420-2025 — MAQUET MAQUET Hello, I own a MAQUET that is covered by recall Z-2420-2025 from MAQUET. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.