ICU Medical recalls ChemoLock Port REF: CL2100 with potential port weld leaks. Affected units have specific UDI and lot numbers. Stop using immediately to prevent leaks.
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
ICU Medical is recalling ChemoLock Port REF: CL2100 devices because the port weld might separate or break during use, causing leaks. This could lead to medication exposure or contamination risks for patients. Affected units have specific UDI and lot numbers.
The port weld in the ChemoLock Port device may separate or break during use, which can cause medication leaks. This could lead to contamination of the patient's bloodstream or exposure to hazardous chemicals.
Patients using ChemoLock Port REF: CL2100 devices with UDI:10887709057002 and lot numbers 14400191 or 14379145 are most at risk. Healthcare providers who administer chemotherapy using these ports should stop using them immediately.
Check for the product name 'ChemoLock Port REF: CL2100' on the device. Affected units have UDI:10887709057002 and lot numbers 14400191 or 14379145.
Discontinue use of the ChemoLock Port REF: CL2100 device immediately to prevent leaks.
The recalled model is ChemoLock Port REF: CL2100.
Affected units have UDI:10887709057002 and lot numbers 14400191 or 14379145.
Check for the product name 'ChemoLock Port REF: CL2100' and look for UDI:10887709057002 with lot numbers 14400191 or 14379145.
We’ve drafted a message you can send ICU Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-0350-2026 — ICU Medical ICU Medical Hello, I own a ICU Medical that is covered by recall Z-0350-2026 from ICU Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.