GE Medical Systems Israel recalls Starcam/Cam: dual-head cardiac imaging cameras with outdated detector support that could fall, causing injury. Affected units have serial numbers in distribution.
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
GE Medical Systems Israel recalls Starcam/Camstar gamma cameras used for cardiac imaging. These cameras, which were past their service guarantee, might have been moved without proper detector support, risking detector falls that could cause serious injury.
The cameras were past their guaranteed service period and could have been moved without proper detector support. This might cause excessive stress on the detector mounting, leading to a detector falling and causing serious injury.
Healthcare facilities using these cardiac imaging cameras are likely affected. Staff and patients using the cameras are at risk if the detector falls.
The recalled cameras are Starcam or Camstar models designed for cardiac imaging. They have serial numbers listed in the distribution records.
Look for the serial number on the camera to confirm if it's in the distribution list.
GE Medical Systems Israel has become aware that certain cameras past their service guarantee might have been moved without adequate detector support, risking detector falls that could cause serious injury.
Check the camera's serial number to see if it's in the distribution list. If it is, contact GE Medical Systems Israel for further instructions.
The recall record does not specify a remedy or fix.
We’ve drafted a message you can send GE Medical Systems Israel to request your refund or repair — edit it as you like.
Subject: Recall Z-0303-2026 — GE Medical Systems Israel GE Medical Systems Israel Hello, I own a GE Medical Systems Israel that is covered by recall Z-0303-2026 from GE Medical Systems Israel. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.