Medistim recalls MiraQ ultrasound devices with model numbers MQC04001, MQC14001, MQU14011, MQV12011. Printer Support Modules may have insufficient soldering, causing wires to loosen and system shutdowns. Affected units need replacement of the module.
Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced.
Medistim is recalling ultrasound devices with specific model numbers because the Printer Support Module may have insufficient soldering. This can cause wires to loosen and the system to shut down unexpectedly. If you own one of these models, you should have the module replaced to prevent unexpected shutdowns.
The Printer Support Module (PSM) may have insufficient soldering, which can cause the internal wires to loosen. This loosening may lead to the ultrasound system shutting down unexpectedly during use. If the system shuts down while in operation, it could disrupt medical procedures.
Users of Medistim MiraQ ultrasound devices with model numbers MQC04001, MQC14001, MQU14011, or MQV12011 are affected. Medical professionals using these devices for intraoperative procedures are most at risk due to potential system shutdowns during critical operations.
Check your Medistim MiraQ System device for model numbers MQC04001, MQC14001, MQU14011, or MQV12011. The device is an ultrasound system used for intraoperative procedures.
The Printer Support Module needs to be replaced to prevent system shutdowns.
Model numbers MQC04001, MQC14001, MQU14011, and MQV12011 are affected.
The Printer Support Module may have insufficient soldering, causing wires to loosen and the system to shut down unexpectedly during use.
No, but you should replace the Printer Support Module to prevent unexpected shutdowns.
We’ve drafted a message you can send Medistim to request your refund or repair — edit it as you like.
Subject: Recall Z-0410-2026 — Medistim Medistim Hello, I own a Medistim that is covered by recall Z-0410-2026 from Medistim. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.