GE Healthcare Finland recalls CARESCAPE ONE monitors with specific battery models (serial numbers listed) due to risk of losing patient monitoring during power failure. Affected units may require battery replacement to prevent delays in recognizing patient status changes.
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
GE Healthcare Finland is recalling CARESCAPE ONE monitors with certain battery models (serial numbers listed) because the batteries can lose their ability to hold charge and provide backup power. If this happens, the monitor may lose patient monitoring capabilities, potentially delaying critical treatment for patients.
The batteries can lose their ability to hold charge and provide backup power. If this occurs, the monitor may lose patient monitoring capabilities, which could delay recognizing critical patient status changes requiring treatment.
Patients using the CARESCAPE ONE monitor with affected batteries (serial numbers listed) are at risk. Healthcare providers who have installed these batteries as spare parts or shipped with the monitor should check if their units are affected.
Check if your CARESCAPE ONE monitor (model 2087075-300) has the listed serial numbers (SV925270018SA, SV925270014SA, and other serials in the record). Affected units were shipped with potentially affected batteries or installed as spare parts.
Identify your monitor's serial number from the label on the device. Compare it to the list of affected serial numbers provided in the recall notice.
If your monitor has an affected serial number, contact GE Healthcare Finland directly for instructions on battery replacement or other remedies.
The CARESCAPE ONE monitor with model number 2087075-300 is affected.
Check the serial number on your monitor. If it matches any of the listed serial numbers (e.g., SV925270018SA, SV925270014SA), your monitor has affected batteries.
No, but you should check if your monitor has an affected serial number and contact GE Healthcare Finland if it does. The hazard does not cause immediate danger but requires monitoring.
We’ve drafted a message you can send GE Healthcare Finland to request your refund or repair — edit it as you like.
Subject: Recall Z-0454-2026 — GE Healthcare Finland GE Healthcare Finland Hello, I own a GE Healthcare Finland that is covered by recall Z-0454-2026 from GE Healthcare Finland. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.