B Braun recalls Outlook IV sets used with Infusomat pumps due to potential medication backflow and inability to prime. Affected units have specific catalog numbers and expiration dates.
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
B Braun is recalling Outlook IV sets used with Infus: 1. The sets could allow medication to flow backward from secondary to primary IV containers. 2. This could cause medication to not be properly primed, risking incorrect dosing.
The Outlook IV sets can allow medication to flow backward from secondary (piggyback) IV containers into primary IV containers. This can cause medication to not be properly primed, leading to incorrect dosing or treatment.
Healthcare professionals using the Outlook IV sets with B Braun Infusomat pumps are most at risk. Patients receiving IV treatments with these sets may also be affected.
Check for catalog number 354214 on the Outlook IV set. These sets were sold with Infusomat pumps and have an expiration date of 36 months from manufacture.
Look for catalog number 354214 on the Outlook IV set.
The Outlook IV sets can allow medication to flow backward from secondary to primary IV containers and may not prime properly, leading to incorrect dosing.
Check for catalog number 354214 on the Outlook IV set. These sets were sold with Infusomat pumps and have an expiration date of 36 months from manufacture.
The recall does not specify a remedy, so contact B Braun Medical Inc for guidance.
We’ve drafted a message you can send B Braun to request your refund or repair — edit it as you like.
Subject: Recall Z-0716-2026 — B Braun B Braun Hello, I own a B Braun that is covered by recall Z-0716-2026 from B Braun. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.