Olympus recalls 10 Fr. Sheath Model A37004A for endoscopic urology use due to reported ceramic tip breakage. Affected units sold with UDI 04042761023092.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Olympus is recalling its 10 Fr. Sheath Model A37004A for endoscopic urological procedures because ceramic tips have broken. This could cause injury during medical procedures if the sheath fails unexpectedly.
The ceramic tip can break during use, potentially causing injury to the patient or medical staff. This breakage could lead to complications during endoscopic procedures if the sheath fails unexpectedly.
Urology doctors and medical staff using the Olympus Sheath Model A37004A for endoscopic procedures are most at risk. Patients undergoing these procedures may also be affected if the sheath breaks during treatment.
Check for the model number A37004A on the sheath. The product was sold with UDI 04042761023092 and all lots are affected.
Inspect the ceramic tip for cracks or signs of breakage before use.
Reach out to Olympus Corporation of the Americas for further guidance on safe use.
The recall indicates reported breakage, but the sheath may still be safe if the ceramic tip is intact. Always check for damage.
Look for model number A37004A and UDI 04042761023092 on the sheath.
The recall does not require immediate stoppage; check for damage and contact Olympus for guidance.
We’ve drafted a message you can send Olympus to request your refund or repair — edit it as you like.
Subject: Recall Z-1454-2026 — Olympus Olympus Hello, I own a Olympus that is covered by recall Z-1454-2026 from Olympus. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.
+ 1 more story