Medline recalls semi-electric lightweight homecare beds with model MDR107002L due to rare fire risk from hand pendant overheating. 58 complaints reported.
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Medline is recalling semi-electric lightweight homecare beds with model number MDR107002L because the hand control pendant can overheat and potentially cause fire if the bed is used beyond its weight limit or encounters obstructions. This risk is rare but serious enough to warrant a recall.
The hand control pendant and its wires can overheat when the bed is operated above its weight limit, encounters obstructions, or has jammed components. In rare cases, this overheating may cause the pendant to spark, burn, melt, smoke, or start a fire.
Homecare users who own the recalled beds with model MDR107002L are most at risk. The risk is higher for users who exceed the bed's weight limit or have obstructions during operation.
Check for beds labeled 'Semi Electric Lightweight Homecare Bed' with model number MDR107002L. All lots are affected, so any bed with this model number should be considered recalled.
Identify the model number on the bed's label to confirm if it matches MDR107002L.
If the bed is used beyond its weight limit, stop using it immediately to prevent overheating risks.
Make sure there are no obstructions blocking bed movement to avoid jammed components.
No, only beds labeled as 'Semi Electric Lightweight Homecare Bed' with model number MDR107002L are affected.
Medline has received 58 complaints related to pendant sparking, burning, melting, smoking, or fire.
If the pendant overheats, stop using the bed immediately and contact Medline for assistance.
We’ve drafted a message you can send Medline to request your refund or repair — edit it as you like.
Subject: Recall Z-1012-2026 — Medline Medline Hello, I own a Medline that is covered by recall Z-1012-2026 from Medline. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.