NIPRO recalls specific conductivity solutions that may cause inaccurate dialysate calibration readings, potentially leading to improper treatment. Affected units include model numbers 10001, 10011, 10021, 10032, 10040, 10040c, and 10041.
due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.
NIPRO recalls certain conductivity solutions used to calibrate dialysis meters because they may report incorrect conductivity values. This could lead to improper dialysis treatment if the calibration is inaccurate.
These solutions are used to calibrate conductivity meters in dialysis systems. If the conductivity values are incorrect, dialysis treatment may be improperly calibrated, potentially leading to ineffective treatment. However, the defect does not cause immediate physical harm.
Healthcare professionals using dialysis equipment who rely on these conductivity solutions for calibration. The risk is low as the defect only affects calibration accuracy, not direct safety.
Check for model numbers 10001, 10011, 10021, 10032, 10040, 10040c, or 10041 on the product label. The solutions were sold with specific lot numbers and UDI codes as listed in the recall notice.
Look for model numbers 10001, 10011, 10021, 10032, 10040, 10040c, or 10041 on the solution container.
The recall is due to a defect in the conductivity values of these solutions, which can result in incorrect calibration readings for dialysis meters.
No, but you should check if you have the affected model numbers and contact NIPRO if you need further assistance.
The recall does not require immediate action, but you should verify the model numbers and contact NIPRO for guidance on next steps.
We’ve drafted a message you can send NIPRO to request your refund or repair — edit it as you like.
Subject: Recall Z-1185-2026 — NIPRO NIPRO Hello, I own a NIPRO that is covered by recall Z-1185-2026 from NIPRO. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.