Philips recalls Allura Xper FD10/10 systems with model 722005 due to potential coolant contact with electrical components if drip tray is not installed, risking system shutdown.
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Philips is recalling Allura Xper FD10/10 systems with model number 722005 because the drip tray under the cooling unit may not be properly installed. If the drip tray is missing, coolant could contact electrical parts, causing short circuits and system shutdowns.
The drip tray beneath the cooling unit may not be installed properly in some systems. If missing, coolant liquid could contact electrical components, causing short circuits that might shut down the system.
Medical professionals using Philips Allura Xper FD10/10 systems with model number 722005 are most at risk if the drip tray is not installed correctly.
Check for model number 722005 on the system. The recall affects specific units with serial numbers listed in the official notice.
Ensure the drip tray is properly installed under the cooling unit. If it's missing or not documented, contact Philips for guidance.
Philips identified that in some units the drip tray beneath the cooling unit was not installed correctly or documentation was inadequate, risking coolant contact with electrical parts.
Coolant liquid could contact electrical components, causing short circuits that might shut down the system.
Check for model number 722005 on your Allura Xper FD10/10 system. The recall includes specific serial numbers listed in the official notice.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-1065-2026 — Philips Philips Hello, I own a Philips that is covered by recall Z-1065-2026 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.