ICU Medical recalls Plum Solo Precision IV Pumps with specific list numbers and software versions due to potential piggyback flush errors that may prevent proper flushing of IV lines.
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
ICU Medical is recalling certain Plum Solo Precision IV pumps that may show error messages when programming piggyback flushes. This could prevent the pump from properly flushing IV lines after treatment, potentially causing medication buildup or incorrect dosing.
The pump's piggyback flush feature can sometimes trigger an error message that prevents the flush from being programmed. This might lead to medication buildup or incorrect dosing if the pump fails to properly clear the IV line after treatment.
Healthcare professionals using ICU Medical Plum Solo Precision IV pumps with list numbers 400010401 (US) or 400011001 (Philippines) and specific software versions are affected. Patients receiving IV treatments with these pumps may be at risk if the flush function fails.
Check for list numbers 400010401 (US) or 400011001 (Philippines) on the pump. Also look for software versions 1.0.2 (US) or 1.0.0 (Philippines).
Verify if the pump displays an Upper Hard Limit or Lower Hard Limit violation message when programming piggyback flushes.
The pump may show error messages that prevent proper piggyback flush programming, which could lead to medication buildup or incorrect dosing.
US models with software version 1.0.2 and Philippines models with software version 1.0.0.
No, but you should check for error messages when programming piggyback flushes.
We’ve drafted a message you can send ICU Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-1120-2026 — ICU Medical ICU Medical Hello, I own a ICU Medical that is covered by recall Z-1120-2026 from ICU Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.