Medline recalls 33 medical procedure kits with Strykeflow 2 Suction Irrigators due to design change causing saline leaks/vapor during use. Affected kits include specific model numbers and lot numbers.
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Medline is recalling 33 different medical procedure kits that contain Strykeflow 2 Suction Irrigators. The design change in 2024 causes irrigation solution to leak or emit vaporized saline during use, which can resemble smoke.
A design change in 2024 allows irrigation solution to travel to the handpiece and battery pack during use. This can cause leaks or vaporized saline (which may look like smoke) to be emitted, potentially causing confusion or safety issues during procedures.
Healthcare professionals using these kits for medical procedures are most at risk. The recall affects specific medical procedure kits sold to hospitals and clinics.
Check for kits labeled with specific model numbers like 'WCH BARI-GASTRECTOMY CDS' (REF CDS860237AF) or 'ROBOTIC PROCEDURE' (REF CDS982128Q). Also look for lot numbers from the list provided in the recall notice.
Look for model numbers and lot numbers listed in the recall notice to confirm if your kit is affected.
Reach out to Medline directly for assistance with the recall process if you have an affected kit.
A design change in 2024 caused irrigation solution to travel to the handpiece and battery pack during use, leading to leaks or vaporized saline emissions.
Check for specific model numbers and lot numbers listed in the recall notice, such as REF CDS860237AF or lot numbers like 25FBK870.
No, but you should check if your kit is affected and contact Medline if needed.
We’ve drafted a message you can send Medline to request your refund or repair — edit it as you like.
Subject: Recall Z-1087-2026 — Medline Medline Hello, I own a Medline that is covered by recall Z-1087-2026 from Medline. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.