Tornier recalls specific lot of perip screws (DWJ318, Lot AZ482: 5.0mmx18mm labeled but 5.00mmx14mm actual). Mismatch risks surgery delays or revision surgery. No remedy provided.
A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.
Tornier is recalling a specific lot of perip screws (DWJ318, Lot AZ4824290) because they are incorrectly labeled. Using the wrong size screws during surgery could cause delays or require revision surgery. This recall affects surgeons and surgical teams using the product.
The screws are labeled as 5.0mmx18mm but are actually 5.00mmx14mm. If the mismatch is detected during surgery, it may cause a minor delay. If not detected and the incorrect screw is used, it could lead to mechanical failure, loosening, or the need for revision surgery.
Surgeons and surgical teams who use the Tornier Perip Screw (DWJ318, Lot AZ4824290) are affected. The risk is highest for those performing procedures where precise screw sizing is critical.
Check for the catalog number DWJ318 and lot number AZ4824290 on the screw packaging. The screws are labeled as 5.0mmx18mm but are actually 5.00mmx14mm.
Look for the catalog number DWJ318 and lot number AZ4824290 on the screw packaging.
The screws are incorrectly labeled as 5.0mmx18mm but are actually 5.00mmx14mm. This mismatch could cause surgery delays or require revision surgery if not detected.
The recall record does not specify a remedy. Check the product label for the catalog number DWJ318 and lot number AZ4824290.
The recall states the issue may cause minor delays or increase risk of revision surgery, but does not mention serious injury risks.
We’ve drafted a message you can send Tornier to request your refund or repair — edit it as you like.
Subject: Recall Z-1237-2026 — Tornier Tornier Hello, I own a Tornier that is covered by recall Z-1237-2026 from Tornier. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.