Medline recalls 24 convenience kits for potential sterility issues due to calibration problems in sterilization equipment. Affected kits include PICC catheter trays and knee arthro basic kits with specific model numbers and lot dates.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline is recalling 24 different convenience kits that may have been affected by calibration issues in sterilization equipment. This could compromise the sterility of the kits, posing a risk of infection during medical procedures.
The calibration issues in the sterilization equipment used to package these kits could lead to a failure in maintaining sterility. This means the kits might not be sterile when used, increasing the risk of infection during medical procedures.
Healthcare professionals using these kits in hospitals or clinics are most at risk. Patients receiving procedures using these kits could potentially be exposed to infection.
Check for model numbers: CVI4300, DYKM1977C, DYKMBNDL130, DYKMBNDL130A, DYKMBNDL4, or LC295. Look for specific lot numbers and dates listed in the recall notice.
Identify the model number and lot number on your kit to confirm if it's affected by this recall.
Medline identified calibration issues in the sterilization equipment used to package these kits, which could impact the sterility assurance level.
Check the model number and lot number on your kit to see if it's affected. If it is, contact Medline for further instructions.
The recall indicates a potential risk to sterility, but the kits are not necessarily unsafe. Affected kits may need to be checked or replaced.
We’ve drafted a message you can send Medline to request your refund or repair — edit it as you like.
Subject: Recall Z-2118-2026 — Medline Medline Hello, I own a Medline that is covered by recall Z-2118-2026 from Medline. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.