Waldemar Link recalls Endo-Model SL Connection Components with serial numbers 250925/0480-0504 and others due to potential plateau screw detachment causing implant failure.
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Waldemar Link is recalling Endo-Model SL Connection Components with specific serial numbers because a plateau screw could detach from the screw shaft, potentially causing implant failure. This recall affects medical devices used in certain surgical procedures.
The plateau screw in the component could detach from the screw shaft due to a longitudinal fracture. This detachment might lead to implant failure, which could cause complications during or after surgery.
Medical professionals who use these components in surgical procedures are likely affected. Patients who have received implants with these serial numbers may also be at risk.
Check for serial numbers starting with 250925 followed by numbers like 0480 through 0504. The product includes a PE-Plateau and is part of the Endo-Model SL Connection system.
Immediately stop using the Endo-Model SL Connection Component if it has a serial number listed in the recall.
The plateau screw could detach from the screw shaft due to a longitudinal fracture, potentially causing implant failure.
Serial numbers starting with 250925 and ending with numbers like 0480, 0481, 0482, 0483, 0484, 0486, 0489, 0491, 0492, 0494, 0503, 0504, 0345, 0346, 0349, 0350, 0351, 0352, 0356, 0369, 0501, 0365, 0368, 0490, 0488, 0479, 0495, 0500, 0496, 0499, 0367.
The recall record does not specify a remedy or replacement option.
We’ve drafted a message you can send Waldemar Link to request your refund or repair — edit it as you like.
Subject: Recall Z-1514-2026 — Waldemar Link Waldemar Link Hello, I own a Waldemar Link that is covered by recall Z-1514-2026 from Waldemar Link. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.