Agiliti Health recalls Adapt Pump (model 61600200) with software version 1.3.0 made before Dec 2025. The pump stops Microclimate Management after 10 minutes of Autofirm use, potentially causing uneven pressure. Affected users should contact Agiliti for a fix.
Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.
Agiliti Health is recalling Adapt Pumps (model 61600200) made before December 2025 that stop the Microclimate Management function after 10 minutes of Autofirm use. This could lead to uneven pressure distribution on the user's body, increasing the risk of discomfort or injury.
The pump stops the Microclimate Management function after 10 minutes of Autofirm use, which can cause uneven pressure distribution. This may lead to discomfort or injury in users who rely on consistent pressure for medical support.
People using Agiliti Adapt Pumps (model 61600200) with software version 1.3.0 made before December 2025 are affected. Those with medical conditions requiring precise pressure management are at higher risk.
Check for the model number 61600200 on the pump. The pump was made before December 4, 2025, and has software version 1.3.0. It comes with Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, or Adapt Convertible EX support surfaces.
Identify the model number on the pump, which should be 61600200 for affected units.
Reach out to Agiliti Health for a software update or replacement to fix the Microclimate Management issue.
No, only Adapt Pumps with model number 61600200 and software version 1.3.0 made before December 4, 2025.
Contact Agiliti Health for a software update or replacement to fix the Microclimate Management issue.
The recall addresses a potential issue with uneven pressure distribution, but the official notice does not specify injury risks.
We’ve drafted a message you can send Agiliti Health to request your refund or repair — edit it as you like.
Subject: Recall Z-1472-2026 — Agiliti Health Agiliti Health Hello, I own a Agiliti Health that is covered by recall Z-1472-2026 from Agiliti Health. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.