Philips recalls Allura Xper FD20/15 X-ray systems with specific system codes due to potential failure to initiate X-ray imaging when using the wired foot switch.
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Philips is recalling all Allura Xper FD20/15 X-ray systems with system code 722058 because the X-ray imaging may not start or may stop unexpectedly when using the wired foot switch. This could lead to incomplete medical imaging procedures.
The X-ray imaging may not initiate or may occur intermittently when using the wired foot switch. This could cause incomplete medical imaging procedures, potentially leading to the need for repeat scans or incorrect diagnoses.
Healthcare professionals using the Allura Xper FD20/15 X-ray system with system code 722058 are most at risk. Patients undergoing X-ray procedures with this specific model may experience incomplete imaging.
Check for system code 722058 on the device. The system code is listed in the product description and can be found on the device's label or documentation.
Verify the system code on the device label or documentation to confirm if it matches 722058.
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Check for system code 722058 on the device label or documentation.
The recall does not specify a remedy, so contact Philips directly for further instructions.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-1671-2026 — Philips Philips Hello, I own a Philips that is covered by recall Z-1671-2026 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.