Depuy recalls ATTUNE Revision Hinge Femoral LT SZ 4 CEM parts with potential sterility compromise due to packaging adhesion. Affected units have specific part numbers and expiration dates.
The external sterile packaging was found adhered to the internal sterile packaging, potentially compromising sterility.
Depuy is recalling a specific hip implant part because the sterile packaging might stick to the inside packaging, risking infection during surgery. This is a serious issue for patients needing hip replacements.
The external sterile packaging can stick to the internal sterile packaging, which might compromise sterility during surgery. This could lead to infection if the implant is not properly sterile.
Patients who received this specific hip implant part during surgery are affected. Surgeons and hospitals using the part are most at risk.
Check for part number 150450104 on the packaging. The expiration date is November 30, 2035, and the lot number is 1007844.
Look for part number 150450104, lot number 1007844, and expiration date November 30, 2035 on the packaging.
Yes, this is a hip implant part for revision surgeries.
Check for part number 150450104, lot number 1007844, and expiration date November 30, 2035 on the packaging.
The recall does not specify a remedy, so contact Depuy for guidance.
We’ve drafted a message you can send Depuy to request your refund or repair — edit it as you like.
Subject: Recall Z-1985-2026 — Depuy Depuy Hello, I own a Depuy that is covered by recall Z-1985-2026 from Depuy. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.