Becton Dickinson recalls Swan-Ganz Jr Catheters (models SGPT54, SGPT64P, SGPT755P) due to leaking hubs that could cause infection or blood loss. Affected units have specific serial numbers and expiration dates.
Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
Becton Dickinson is recalling Swan-Ganz Jr Catheters with models SGPT54, SGPT64P, and SGPT755P because the blue proximal injectate lumen hub may leak or break, leading to infection, medication loss, or blood loss. This risk affects medical devices used in critical care settings.
The catheter's blue proximal injectate lumen hub can leak or break due to manufacturing material changes. This may cause infection, medication loss, or blood loss in patients receiving critical care.
Healthcare professionals using Swan-Ganz Jr Catheters for critical care patients are most at risk. Patients receiving these catheters through medical facilities may also be affected.
Check for the model numbers SGPT54, SGPT64P, or SGPT755P on the catheter. Affected units have specific serial numbers and expiration dates listed in the recall notice.
Identify the model number on the catheter to see if it matches SGPT54, SGPT64P, or SGPT755P.
Reach out to Becton Dickinson for further instructions on handling the recalled catheters.
The blue proximal injectate lumen hub may leak or break due to manufacturing material changes, leading to infection, medication loss, or blood loss.
Models SGPT54, SGPT64P, and SGPT755P are affected by this recall.
Check the model number on the catheter and compare it to SGPT54, SGPT64P, or SGPT755P. Affected units have specific serial numbers and expiration dates listed in the recall notice.
We’ve drafted a message you can send Becton Dickinson to request your refund or repair — edit it as you like.
Subject: Recall Z-2203-2026 — Becton Dickinson Becton Dickinson Hello, I own a Becton Dickinson that is covered by recall Z-2203-2026 from Becton Dickinson. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.