D.O.R.C. recalls TDC VELOCE 27G cutters with loose outer knives. Affected units have specific article numbers and lot numbers. Stop using if knife becomes loose during use.
An increase in incidence in complaints of a loose outer knife and reports of the outer knife becoming loose during use.
D.O.R.C. is recalling TDC VELOCE 27G Two Dimensional Cutters due to reports of the outer knife becoming loose during use. This could lead to injury if the knife slips while cutting. Affected units have specific article numbers and lot numbers listed in the recall.
The outer knife in the cutter can become loose during use, which may cause it to slip and potentially injure the user. This is a mechanical issue that could happen with normal use, especially during precise eye procedures.
Users of the TDC VELOCE 27G Two Dimensional Cutter are affected. Medical professionals who use this device for eye procedures are most at risk due to potential injury from a loose knife.
Check for article numbers: 9268.VIT27, 9311.27G01, 9311.27G02, 9311.27K01, 9311.27K02, 9311.27K03, 9627.G0101, 9627.G0201, 9311.27G01-00. Also look for lot numbers starting with 202511001, 202601070, or patterns like 51682-*-*-1.
If the outer knife feels loose or moves during use, stop using the device immediately.
D.O.R.C. is recalling TDC VELOCE 27G Two Dimensional Cutters due to reports of the outer knife becoming loose during use.
Check for article numbers: 9268.VIT27, 9311.27G01, 9311.27G02, 9311.27K01, 9311.27K02, 9311.27K03, 9627.G0101, 9627.G0201, 9311.27G01-00. Also look for lot numbers starting with 202511001, 202601070, or patterns like 51682-*-*-1.
If the outer knife feels loose or moves during use, stop using it immediately. Contact D.O.R.C. for further instructions.
We’ve drafted a message you can send D.O.R.C. to request your refund or repair — edit it as you like.
Subject: Recall Z-2324-2026 — D.O.R.C. D.O.R.C. Hello, I own a D.O.R.C. that is covered by recall Z-2324-2026 from D.O.R.C.. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.