Medline recalls oral restoration medical kits with irregular catheter surfaces and detachment risks. Affected kits have specific model numbers and lot codes.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Medline is recalling oral restoration medical procedure kits due to potential catheter surface issues and detachment problems. These flaws could lead to improper use during medical procedures.
The catheter surface finish may be irregular, causing detachment from the funnel. This could lead to improper use during medical procedures if the catheter doesn't meet specifications.
Healthcare providers using Medline's oral restoration kits are most at risk. The recall affects kits sold with specific model numbers and lot codes.
Check for Medline Kit Number SKUDYNJ86253A on the kit label. Lot numbers 25KBC063 or 25JBO955 are also associated with this recall.
Look for Medline Kit Number SKUDYNJ86253A and lot numbers 25KBC063 or 25JBO955.
Medline identified issues with catheter surface finish and detachment problems during internal reviews.
Check the kit label for the model number and lot numbers to confirm if it's affected.
The recall addresses potential issues with catheter surface finish and detachment, but no immediate safety risks are described.
We’ve drafted a message you can send Medline to request your refund or repair — edit it as you like.
Subject: Recall Z-2490-2026 — Medline Medline Hello, I own a Medline that is covered by recall Z-2490-2026 from Medline. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.