Medline recalls convenience kits with Swan-Ganz catheters linked to leaks and breakage, risking infection and blood loss. Affected kits include specific model numbers and lots.
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Medline is recalling convenience kits that contain Swan-Ganz catheters known to leak or break, which could lead to infections, medication loss, or blood loss. This recall affects specific model numbers and production lots.
The catheters have a blue proximal injectate lumen hub that can leak or break, causing potential infections, medication loss, or blood loss. This issue occurs during normal use in medical settings.
Healthcare professionals using Medline convenience kits with the recalled Swan-Ganz catheters are most at risk. Patients receiving care from these kits could also be affected.
Check for model numbers DYNJ901328G, DYNJ903465K, DYNJ909090G, DYNJ906102M, or DYNJ902664M/N on the kits. The kits were sold with specific production lots and UDI-DI numbers listed in the recall notice.
Identify the model number on the kit to see if it matches the recalled ones (DYNJ901328G, DYNJ903465K, etc.).
Reach out to Medline directly for guidance on next steps if the kit is affected.
Becton Dickinson (BD) has received complaints of leaks and breakage in the Swan-Ganz catheters, which can lead to infections, medication loss, or blood loss.
The affected model numbers include DYNJ901328G, DYNJ903465K, DYNJ909090G, DYNJ906102M, and DYNJ902664M/N.
Check the model number on the kit; if it matches the listed numbers (DYNJ901328G, DYNJ903465K, etc.), it is affected.
We’ve drafted a message you can send Medline to request your refund or repair — edit it as you like.
Subject: Recall Z-2607-2026 — Medline Medline Hello, I own a Medline that is covered by recall Z-2607-2026 from Medline. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.