Product Recall Number: 12742
Why was Ondal AC2000 Television Mounting System recalled?
The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.
Product Recall Number: 12742
Product Safety Warning Number:
Date: 2012-07-17
Product Safety Warning Date:
Name Of Product: Ondal AC2000 Television Mounting System
Description: The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.
Hazard Description: The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.
Consumer Action: None Listed
Original Product Safety Warning Announcement: None Listed
Remedy Type: Repair
Units: 215 distributed to 5 hospitals
Incidents: The firm received one report of a mounting system falling while being positioned. No injuries have been reported.
Remedy: Secure the unit to prevent movement and contact TRUMPF Medical Systems for information on receiving a repair kit that includes the missing washers.
Sold At: TRUMPF Medical Systems Inc. nationwide From January 2005 through December 2011 for about $2,100.
Importers: Ondal Medical Systems of America Inc., of Sandston, VA and TRUMPF Medical Systems Inc., of Charleston, SC
Manufacturers: None Listed
Distributors: TRUMPF Medical Systems Inc., of Charleston, SC
Manufactured In: Germany
Custom Label: None Listed
Custom Field: None Listed