Every recall we track for Haemonetics Corporation — 29 in all — plus the complaints and injury reports the public filed about them.
Haemonetics recalls TEG 5000 Functional Fibrinogen Reagent (Item 07-034) due to potential misdiagnosis from outdated blood reference ranges. All lots affected.
Haemonetics recalls TEG5000 analyzers with specific software versions that may display incorrect PlateletMapping ADP or AA results. Affected users should update software to fix the issue.
Haemonetics recalls TEG Manager software versions 1.1.0, 3.0.0, 4.0.0, 4.1.0, and 4.1.1 used with TEG 5000. These versions display out-of-range alerts incorrectly due to rounding errors in reference values.
Haemonetics recalls Japan Low Vol 125ml CS5 Set (Lot 118011) due to bowl cracks causing leaks. This may lead to device errors and incomplete blood washing. Affected units were sold between Oct 2014 and Oct 2019.
Haemonetics recalls Japan Basic Hi Speed CS5 Set (LN 263J) due to potential leaks in bowl sets that could cause device errors and incomplete washing. Affected units were sold between Oct 2014 and Oct 2019.
Haemonetics recalls Japan Cell Saver Elite Set 125ml products due to potential leaks in inner core causing device errors. Affected units were sold between October 2014 and October 2019. No remedy specified in recall notice.
Haemonetics recalls certain Cell Saver bowl sets that may leak fluid, causing device errors and incomplete washing. Affected units were sold between October 2014 and October 2019.
Haemonetics recalls CS Elite Processing Set (125ml) autotransfusion devices with bowl cracks or missing crack-check instructions due to potential leaks causing device errors. Affected units sold between Oct 2014 and Oct 2019.
Haemonetics recalls CS5/5+ Fastpacks with bowl cracks that may cause leaks and device errors. Affected units were sold between Oct 2014 and Oct 2019. Check for cracks in the bowl.
Haemonetics recalls Japan DISP SET models 263J, 205J, and 208 due to potential leaks in bowl sets that may cause device errors. Affected units were sold between October 2014 and October 2019.
Haemonetics recalls JAPAN CS Elite Fastpack 225ml units with bowl cracks that may leak blood during blood recovery procedures. Affected units have specific lot numbers from October 2014 to 2019.
Haemonetics recalls CS5/5+ Fastpacks with bowl cracks that may cause leaks and device errors. Affected units include specific lots from Oct 2014 to Oct 2019.
Haemonetics recalls CS5/5+ Bowl Kit-125 mL Autotransfusion Devices with cracks or missing instructions that may cause leaks and device errors. Affected units were sold between Oct 2014 and Oct 2019.
Haemonetics recalls CS Elite Fastpack 225mL autotransfusion devices with potential leaks causing fluid trapping and device errors. Affected units have bowl cracks or missing crack-check instructions from Oct 2014 to Oct 2019.
Haemonetics recalls CS Elite Processing Set (225 ml) autotransfusion devices with potential leaks in bowl sets that may cause incomplete washing and device errors.
Haemonetics recalls Japan CS Elite FASTPACK 125ml sets with bowl cracks or missing crack-check instructions. Leaks may cause device errors and incomplete washing. Affected units sold between Oct 2014 and Oct 2019.
Haemonetics recalls Japan Disp Set models 261J, 205J, 208 due to potential leaks from bowl sets causing device errors. Affected units may have cracks or missing instructions for checking cracks.
Haemonetics recalls Cell Saver 5/5+ Bowl Kit (225 ml) autotransfusion devices with bowl cracks or missing crack-check instructions. Leaks may cause fluid trapping and device errors during blood processing.
Haemonetics recalls CS Elite Fastpack 125mL autotransfusion devices with potential leaks causing fluid trapping and device errors. Affected units have bowl cracks or missing crack-check instructions from Oct 2014 to Oct 2019.
Haemonetics recalls CS5/5+ Fastpacks (125mL) from Oct 2014-2019 due to potential leaks causing device errors. Affected units may need repair or replacement.
Haemonetics recalls Japan DISP SET models 263J and 205J (2014-2019) due to potential leaks in bowl sets causing device errors. Affected units may need replacement.
Haemonetics recalls CS Elite Fastpack 225mL blood recovery devices with potential leaks causing fluid traps and error messages. Affected units have specific lot numbers from Oct 2014 to Oct 2019.
Haemonetics recalls Japan Disp Set models 261J and 205J due to potential leaks in blood processing units that could cause device errors and incomplete washing. Affected units may have bowl cracks or missing instructions for crack checks.
Haemonetics recalls CS5/5+ Fastpacks with bowl cracks or missing crack checks. Leaks may cause device errors and incomplete blood washing. Affected units sold between Oct 2014-2019.
Haemonetics recalls Japan Cell Saver Elite Set (225ml) due to potential leaks from bowl cracks causing device errors. Affected units include specific lots manufactured between Oct 2014 and Oct 2019.
Haemonetics recalls OrthoPAT QuickConnect Reservoirs (OPT-R-1000) due to missing cannula that prevents blood from emptying. Affected units may not process collected blood safely during surgery.
Haemonetics recalls OrthoPAT System-Integrated Set (Product Code 1150H-00) due to missing cannula in reservoirs, preventing blood from being processed. Affected units sold with Lot #1017049 and 1150H-00.