Every recall we track for Getinge Disinfection Ab — 5 in all — plus the complaints and injury reports the public filed about them.
Getinge recalls 88-Series washer-disinfectors with model 88-5 due to overheating risk if circulation pressure calibration is incorrect. Affected units may cause serious injury. Contact Getinge for remedy.
Getinge recalls Aquadis 56 washer disinfectors with model 56A due to pump seal wear from improper frequency inverter programming. Affected units have specific serial numbers. Contact Getinge for details.
Getinge recalls 56M model Washer Disinfector Aquadis 56 units with faulty frequency inverters that cause premature axial seal wear. Affected units have specific serial numbers listed in the recall notice.
Getinge recalls AGS, FSLC, FSUC, RC accessories for 88 Series Washer-Disinfectors due to potential cart unloading without trolley docking, risking falls and damage.
Getinge recalls accessories for 86 Series Washer Disinfector models due to potential cart unloading without trolley docking, risking falls and damage. Affected units have specific serial numbers.